User Tools

Site Tools


trekstart

TrakStart Elecromagnetic Tracker

Installation

  1. Installation package is not signed, so Windows SmartScreen will show a WARNING. Just click on More Info → Run Anyway
  2. No other special actions required for the installation

Configuration

  1. With the device connected and turned on, open Cubes.exe
  2. Click Run
  3. The attached sensors will appear along with and Interference Score. For high accuracy. this score should not be higher than 2

Troubleshooting

Q: Interference score is higher than 2

A: There are some things you can do:

  • Usually metal messes up the electromagnetic field. Place the entire setup on a carton or wood board
  • Remove as much as possible all nearby metal and electronic devices
  • Move the tracker closer to the sensors
trekstart.txt · Last modified: 2024/01/18 15:44 by costin